抱歉,此产品系列已不再提供。
PD300-CIE是一款配备旋转座的光度传感器。具有一个眼响应曲线传感器,以勒克斯(Lux)为单位进行测量。其测量范围为20mLux - 200kLux。该传感器配备一根1.5米电缆,用于连接至一个仪表或PC接口。
- Eye response curve sensor with units of Lux
- 20 mLux to 200 kLux range
- Swivel mount 查看所有功能
规格
- 孔径尺寸2.4 x 2.8 mm
- 材料Silicon
- 光谱范围400-700 nm
- 功率范围20 mLux to 200 kLux
- 最大平均功率密度10 W/cm²
- 最大脉冲能量1 µJ
- 响应时间0.2 s
- 电缆长度1.5 m
- 尺寸118 x 10 x 18 mm (LxWxD)
- CE 符合性Yes
- UKCA 符合性Yes
- 中国 RoHS 符合性Yes
功能
Photometric Measurements: Measuring Illuminance with PD300-CIE
In applications where a human observer is involved (for example illumination applications), it is often important to measure using the eye-response-matched Photometric system of units. Ophir’s PD300-CIE is a photometric sensor, and is designed to measure illuminance (in units of Lux or Foot-Candles).
Measuring Very Low Laser Powers with a Photodiode Sensor
In this short “Basics” video, we review in general the use of photodiode sensors for measuring very low laser powers.
常见问题
- Do I need to recalibrate my instrument? How often must it be recalibrated?答案
Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.
- Can I use a sensor from the PD300 family to measure average power of pulsed lasers?答案
In general yes, but several technical issues need to be kept in mind (most of which are results of the fast physical response time of these sensors):
- The pulse rate should be more than about 30Hz, otherwise the reading is unstable. At higher pulse frequencies, the sensor will respond as if the beam were CW.
- It is possible for a pulsed beam to have average power within the sensor spec and yet have the energy of the pulses themselves be high enough to cause a momentary saturation of the sensor. It is important to be sure that pulse energy is also within sensor spec (the parameter "Max pulse energy" is included in all specs for the PD300 family, for just this reason).
- The beam diameter should be no less than about 1mm .
- The average power and power density restriction in the spec should not be exceeded
Note: At the maximum pulse energy limit given in the spec, the reading will be saturated by about 5%, i.e. the reading will be about 5% lower than it should be. At 1/3 the maximum, the saturation will be about 1%.
配件
Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
| 对比 | 型号 | 图纸、CAD和参数 | 库存 | 价格 | ||
|---|---|---|---|---|---|---|
![]() | 7E01125AOptical Sensor Cable, 5 m Length | |||||
![]() | 7E01125BOptical Sensor Cable, 10 m Length | |||||
![]() | XWAR-SENSORExtended Warranty for Sensor |
资源
Data Sheets
Catalogs
Power Sensors Catalog(11.2 MB, PDF) Laser Power & Energy Measurement and Laser Beam Analysis Catalog(27.5 MB, PDF)
Drawings & CAD
PD300-CIE Drawing(117.3 kB, PDF)
Videos
Technical Notes
5 Situations Where Laser Performance Measurement is Necessary Effect of Ambient Conditions on Laser Measurements How to Properly Select a Laser Power or Energy Sensor