PD300是一款通用型光电二极管激光测量传感器,配备旋转座及可拆卸式滤光片,具有10x10mm孔径。未装配滤光片时,其光谱范围为350 - 1100nm,功率测量范围为500pW - 30mW。装配滤光片时,其光谱范围为430nm - 1100nm,功率测量范围为200µW - 300mW。该传感器具有Ophir专有的自动背景减除功能。该传感器配备一根1.5米电缆,用于连接至一个表头或PC接口。
- 350 to 1100 nm silicon photodiode sensor
- 500 pW to 300 mW power measurement range
- Swivel mount for hard to measure places
- Comes with filter in / filter out options
- Patented automatic background subtraction
- Fiber optic adapters available 查看所有功能
规格
- 孔径尺寸10 x 10 mm
- 材料Silicon
- 光谱范围Filter in: 430-1100 nm
Filter out: 350-1100 nm - 功率范围Filter in: 2 µW - 300 mW
Filter out: 500 pW - 30 mW - 最大平均功率密度Filter in: 50 W/cm²
Filter out: 10 W/cm² - 最大脉冲能量Filter in: 30 µJ
Filter out: 3 µJ
- 响应时间0.2 s
- 电缆长度1.5 m
- 尺寸120 x 13 x 21 mm (LxWxD)
- CE 符合性Yes
- UKCA 符合性Yes
- 中国 RoHS 符合性Yes
功能
Measuring Very Low Laser Powers with a Photodiode Sensor
In this short “Basics” video, we review in general the use of photodiode sensors for measuring very low laser powers.
Effects of Incidence Angle on Measurements
In this video, you will learn to what degree a beam’s incidence angle matters, and for which sensor types this should be taken into account.
常见问题
- How do I take off the removable filter from the PD300?答案
Removing External Filter from PD300:

Step 1 – Starting position
Step 2
Step 3 - What does the PD300 "Background Subtraction" do?答案
Ophir's Photodiode PD300 and PD300-1W sensors offer automatic background subtraction so the measurement is not sensitive to room light. With "filter out" (i.e. the external filter removed for low light measurements), 2 separate detector elements are visible. The beam to be measured is incident only on the outer of the 2 detectors, but background light reaches both detectors. The instrument will show the power measured by the outer detector minus that measured by the inner detector.This patented method cancels out 95% - 98% of background light under normal room conditions, even if it is constantly changing.
- Do I need to recalibrate my instrument? How often must it be recalibrated?答案
Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.
- Can a lost PD300 filter be replaced?答案
Technically it could be replaced, but it is not just a matter of replacing the filter. Since the PD300 is a "calibrated" sensor it requires that the filter also be "calibrated". Especially since the PD300 response varies with wavelength, it requires that both the PD300 and the filter be calibrated over the entire spectral range with a monochromator. Because of the cost to calibrate the replacement filter with the PD300 sensor, we recommend purchasing a new PD300 sensor when a replacement filter is needed.
- The total error when using a PD-300 head is listed as +/- 3%. Is that 3% of the reading or 3% of the total range?答案
It's +/-3% of the reading from full scale down to 5% of full scale. Below 5% of full scale one should switch to next range down for the best accurate linear results.
- Can I use a PD300 photodiode sensor to measure power of a scanned beam?答案
For measuring power of a scanned beam we recommend using the BC20, and not the PD300. Since a scanned beam will spend only a fraction of the time of each scan on the detector, the average power measured by the detector will correspondingly be only a fraction of the actual power of the beam. The BC20 is specially designed for such applications by having a peak-hold circuit integrated in its electronics.
- Can I use a PD300 sensor inside a thermal chamber, in which the temperature cycles beyond the recommended operating temperature range?答案
It should be okay, as long as:
- the wavelength is not near the long wavelength limit where the PD300 has a large temperature dependence;
- there is no condensation on the window of the detector which could interfere with the beam and affect the reading.
We suggest the customer does an experiment with a stable laser source (such as a pointer laser) shining in through a window onto the detector while the unit is temperature cycled to see if the reading changes. The final measurement should be back at the original temperature so as to make sure the laser hasn’t changed.
- Why do Photodiode-based power sensors like the PD300 have a specified “Max pulse energy”?答案
Although these sensors measure average power (of both CW and repetitively pulsed beams), not pulse energy, it is possible for a pulsed beam to have average power within the sensor’s rated limits and yet have the energy of the pulses themselves be high enough to cause a momentary saturation of the sensor. It is important to be sure that pulse energy is also within sensor spec – not just the average power. This is explained in detail in this White Paper.
- Can I use a sensor from the PD300 family to measure average power of pulsed lasers?答案
In general yes, but several technical issues need to be kept in mind (most of which are results of the fast physical response time of these sensors):
- The pulse rate should be more than about 30Hz, otherwise the reading is unstable. At higher pulse frequencies, the sensor will respond as if the beam were CW.
- It is possible for a pulsed beam to have average power within the sensor spec and yet have the energy of the pulses themselves be high enough to cause a momentary saturation of the sensor. It is important to be sure that pulse energy is also within sensor spec (the parameter "Max pulse energy" is included in all specs for the PD300 family, for just this reason).
- The beam diameter should be no less than about 1mm .
- The average power and power density restriction in the spec should not be exceeded
Note: At the maximum pulse energy limit given in the spec, the reading will be saturated by about 5%, i.e. the reading will be about 5% lower than it should be. At 1/3 the maximum, the saturation will be about 1%.
配件
Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
| 对比 | 型号 | 图纸、CAD和参数 | 库存 | 价格 | ||
|---|---|---|---|---|---|---|
![]() | 7Z02418CDRH Adapter, Ø7 mm Aperture, Wand Type Power Sensor | 有库存 | ||||
![]() | 7Z08336CDRH Adapter, Ø3.5 mm Aperture, Wand Type Power Sensor | 有库存 | ||||
![]() | 7Z08221Fiber Adapter, SC Type | 有库存 | ||||
![]() | 7Z02210Fiber Adapter, ST Type | 有库存 | ||||
![]() | 7Z02213Fiber Adapter, FC Type | 有库存 | ||||
![]() | 7Z02212Fiber Adapter, SMA Type | 有库存 | ||||
![]() | 7E01125AOptical Sensor Cable, 5 m Length | |||||
![]() | 7E01125BOptical Sensor Cable, 10 m Length | |||||
![]() | 7E01125COptical Sensor Cable, 3 m Length | |||||
![]() | XWAR-SENSORExtended Warranty for Sensor |
资源
Data Sheets
Standard Photodiode Sensors 50pW to 3W Datasheet(394.6 kB, PDF)
Catalogs
Power Sensors Catalog(11.2 MB, PDF) Laser Power & Energy Measurement and Laser Beam Analysis Catalog(27.5 MB, PDF)
Drawings & CAD
PD300 Drawing(151.5 kB, PDF)
Videos
Application Notes
Technical Articles
PD300 Photodiode Sensor Spectral Response Ophir Power/Energy Meter Calibration Procedure and Traceability/Error Analysis Laser Measurements in Materials Processing LIDAR Guns, Accuracy, and Speeding Tickets Low Frequency Power Mode
Technical Notes
5 Situations Where Laser Performance Measurement is Necessary Effect of Ambient Conditions on Laser Measurements How do I know what range, or scale, to set my power/energy meter to? Measuring Average Power of Pulsed Lasers with Photodiodes(114.5 kB, PDF)