PD300-IRG是一款针对电信应用的灵敏性光电二极管激光测量传感器。该传感器可接受光纤输入或自由空间,具有5mm孔径。配备可拆卸式滤光片。未装配滤光片时,其光谱范围为800 - 1700nm,功率测量范围为10pW - 800µW。装配滤光片时,其光谱范围为950 - 1700nm,功率测量范围为20µW - 150mW。该传感器配备一根1.5米电缆,用于连接至一个表头或PC接口。
- InGaAs photodiode with 800-1700 nm spectral range
- 10 pW to 200 mW power measurement range
- Comes with removable filter
- Ø5 mm aperture
- Fiber optic adapters available 查看所有功能
规格
- 孔径尺寸Ø5 mm
- 材料InGaAs
- 光谱范围Filter in: 950-1630 nm
Filter out: 900-1630 nm - 功率范围Filter in: 1 µW - 200 mW
Filter out: 20 pW - 800 µW - 最大平均功率Filter in: 50 W/cm²
Filter out: 5 W/cm² - 最小脉冲能量Filter in: 100 µJ
Filter out: 1 µJ
- 响应时间0.2 s
- 电缆长度1.5 m
- 尺寸Ø30 x 29 mm
- CE 符合性Yes
- UKCA 符合性Yes
- 中国 RoHS 符合性Yes
功能
Measuring Very Low Laser Powers with a Photodiode Sensor
In this short “Basics” video, we review in general the use of photodiode sensors for measuring very low laser powers.
How to Measure Laser Power in Telecom Applications
Optical laser measurements in telecom applications have unique challenges. This video will introduce you to the solutions Ophir offers to help you.
常见问题
- Can I use a PD300 photodiode sensor to measure power of a scanned beam?答案
For measuring power of a scanned beam we recommend using the BC20, and not the PD300. Since a scanned beam will spend only a fraction of the time of each scan on the detector, the average power measured by the detector will correspondingly be only a fraction of the actual power of the beam. The BC20 is specially designed for such applications by having a peak-hold circuit integrated in its electronics.
- The total error when using a PD-300 head is listed as +/- 3%. Is that 3% of the reading or 3% of the total range?答案
It's +/-3% of the reading from full scale down to 5% of full scale. Below 5% of full scale one should switch to next range down for the best accurate linear results.
- When using the fiber optic adaptor, how do we handle power loss due to the fiber relative to calibration?答案
All Ophir power meters, including photodiode power meters, have an air gap between the fiber tip and the sensor. Therefore they measure the power emitted by the fiber into the air and do not take into account any reflection losses there are in the fiber. Therefore, if in actual use, the fiber will be coupled with no loss to another element, then the losses should be added to the reading. These losses are usually about 4%. Thus if the reading on the Ophir meter is say 100 mW, then in lossless use, the real power will be 104 mW.
- Can the PD300 be used with the filter in and with fiber adapter at the same time?答案
Yes, the adapter is arranged so the user can use it either with filter out or in.
- Do I need to recalibrate my instrument? How often must it be recalibrated?答案
Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.
- Can I use a PD300 sensor inside a thermal chamber, in which the temperature cycles beyond the recommended operating temperature range?答案
It should be okay, as long as:
- the wavelength is not near the long wavelength limit where the PD300 has a large temperature dependence;
- there is no condensation on the window of the detector which could interfere with the beam and affect the reading.
We suggest the customer does an experiment with a stable laser source (such as a pointer laser) shining in through a window onto the detector while the unit is temperature cycled to see if the reading changes. The final measurement should be back at the original temperature so as to make sure the laser hasn’t changed.
- When measuring a fiber output, should I put the fiber tip right up against the PD300 detector? If not, how close should I come?答案
The PD300 sensors are not designed to measure with the fiber pushed up right against the detector surface. It may be reading lower in such a case due to saturation of the detector from the concentrated beam or higher due to back reflections off the detector and back again from the fiber tip. The optimal reading will be where the beam is expanded to a size of 2-5mm diameter. Therefore, you should back off the fiber to a distance where the beam has expanded somewhat. Do not back off too far, otherwise if the nominal beam size is larger than given above, you may lose some of the beam off the edges of the detector.
- Why do Photodiode-based power sensors like the PD300 have a specified “Max pulse energy”?答案
Although these sensors measure average power (of both CW and repetitively pulsed beams), not pulse energy, it is possible for a pulsed beam to have average power within the sensor’s rated limits and yet have the energy of the pulses themselves be high enough to cause a momentary saturation of the sensor. It is important to be sure that pulse energy is also within sensor spec – not just the average power. This is explained in detail in this White Paper.
配件
Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
| 对比 | 型号 | 图纸、CAD和参数 | 库存 | 价格 | ||
|---|---|---|---|---|---|---|
![]() | 7Z08216Fiber Adapter, FC Type | 有库存 | ||||
![]() | 7Z08222Fiber Adapter, SMA Type | 有库存 | ||||
![]() | XWAR-SENSORExtended Warranty for Sensor |
资源
Data Sheets
Standard Photodiode Sensors 20pW to 300mW Datasheet(376.8 kB, PDF)
Drawings & CAD
PD300-IRG Drawing(97.5 kB, PDF)
Catalogs
Power Sensors Catalog(11.2 MB, PDF) Laser Power & Energy Measurement and Laser Beam Analysis Catalog(27.5 MB, PDF)
Videos
Application Notes
Technical Articles
PD300 Photodiode Sensor Spectral Response Ophir Power/Energy Meter Calibration Procedure and Traceability/Error Analysis Laser Measurements in Materials Processing Low Frequency Power Mode Measuring Average Power of Pulsed Lasers with Photodiodes
Technical Notes
PD300-IRG-V2 User Notes(264.7 kB, PDF) 5 Situations Where Laser Performance Measurement is Necessary Effect of Ambient Conditions on Laser Measurements How do I know what range, or scale, to set my power/energy meter to? Measuring Average Power of Pulsed Lasers with Photodiodes(114.5 kB, PDF) How to Properly Select a Laser Power or Energy Sensor