Der LBS-400 ist ein optionaler Strahlabschwächer für Pyrocams. Er erlaubt die Messung von Laserstrahlen im UV- oder IR-Bereich mit einem Durchmesser von max. 25,4 mm und Leistungen von 10mW bis ca. 500W(1). Der Strahlabschwächer erhält die ursprüngliche Polarisation des Laserstrahls.
- Measure UV, NIR or IR wavelength laser beams
- Diameters up to 1 in. (25.4 mm)
- Powers from 10 mW to ~500 W Alle Funktionen anzeigen
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | SP90351LBS-400-UVBeam Sampler Attachment, Inconel, 193-355 nm, 10 mW to 500 W | |||||
![]() | SP90354LBS-400-NIRBeam Sampler Attachment, Bulk ND, 1064 nm, 10 mW to 500 W | 1 Week | ||||
![]() | SP90349LBS-400-IRBeam Sampler Attachment, CaF2, 10.6 μm, 10 mW to 500 W | 1 Week |
Spezifikationen
Funktionen
LBS-400 Beam Splitters
The beam sampler operates by reflecting the incoming beam from the front surfaces of a pair of wedges through 90 degrees into the camera. Approximately 99% of the beam is transmitted through each beam sampler with 0.01% passed on to the camera. A set of adjustable filters are provided to make final intensity adjustments to the beam before it reaches the camera imager.
Häufig gestellte Fragen
- What should be my safety concerns while using LBS-400?Antwort
Regular laser safety precautions should apply. Please note, that the main incident beam will exit the LBS-400 refracted approximately 6°. Provision must be made to safely contain the transmitted beam by beam dump or power meter.
- How to deal with removable Beam Blocker?Antwort
The beam blocker is designated to keep one wedge surface protected during handling and shipping. It can be left attached for low power laser operations. However, in the case of high-power laser irradiation scattered light might interfere with beam profile measurement and cause offset. In case of such offset, we recommend removing the beam blocker.
After removal, an additional laser output, having less than 0.1% of incident beam power will be reflected. That beam could be used for simultaneous measurement of laser power via Ophir Power Meter in addition to beam profiler.
- Can I use LBS-400 both for beam profiling and power measurement?Antwort
Yes, the main function of LBS-400 beam splitter is Beam profiling, however, it performs a 3-way beam splitting:
Less than 0.01% (1/103) of the beam is reflected in one direction and less than 1% in another while the remaining 99% of the incident laser beam is transmitted.
This allows user to measure one of the 3 beams, depending on available power measurement equipment and setup convenience.
Ressourcen
Data Sheets
LBS-400 Beam Splitters Datasheet(106 kB, PDF)
Drawings & CAD
LBS-400 Drawing(419.3 kB, PDF)
Technical Notes
LBS-400 User Notes(585.7 kB, PDF) Integrated Solutions