L300W-LP2-50 | Laser Thermal Power Sensors | Power Sensors - Ophir

L300W-LP2-50

7Z02776
描述: 

L300W-LP2-50是一款用于高功率密度和长脉冲激光的水冷热电堆功率/能量激光测量探头。其孔径为50mm,可连续测量0.5W至300W的功率,以及间歇测量200mJ-300J的功率。其高损伤阈值LP2涂层可覆盖0.35 – 2.2以及至10.6μm的光谱范围。探头配有1.5米长的标准电缆,用于连接仪表或PC适配器。

Specification

  • LP2
  • Ø50mm
  • 0.35-2.2µm, 10.6µm
  • 0.5W-300W
  • 200mJ-300J
  • N.A.
  • 90 L x 90 W x 20 D (mm)
  • 300J
  • 12kW/cm²
  • 2.5 s
  • 0.07J/cm²
  • 90J/cm²
  • 300W
  • N.A.
  • CE, China RoHS
Need help finding the right sensor? Try our 传感器查询

FAQ

What are the advantages of the new LP2 absorber coating vs the previous LP1 coating

The new LP2 coating has a number of advantages vs the previous LP1 coating:

 

Absorber LP1 LP2
Damage Threshold at 1kW 6kW/cm² 10kW/cm²
Damage Threshold at 3kW 2.5kW/cm² 5kW/cm²
Damage Threshold at 5kW 2kW/cm² 2.5kW/cm²
Pulsed damage threshold for 10ms pulses 160J/cm² 400J/cm²
Spectral Absorption See graphs below
Angle dependence See graphs below

Ophir High Damage Threshold Absorbers LP2 vs. LP1

Read TutorialRead as PDF

Was this FAQ helpful to you? yes no
Close

I see Ophir has released some new thermal sensors with an absorber called “LP2”. What is it?

The new “LP2” type sensors are specially designed for beams having high power and high power density (and for pulsed beams, high energy density). The LP2 sensors are replacing the equivalent LP1 sensors; as impressive as the LP1 is, the LP2 was developed with the following improvements:

  • Very high damage threshold, for both power density and energy density, for long pulse and CW beams;
  • Spectrally flat; since its absorption remains constant at widely differing wavelengths, this means that sensors based on the LP2 can be used for "white light" or polychromatic beams;
  • Very high level of absorption (as high as 96%, depending on wavelength), meaning much less light is scattered back, which for high power beams is an important benefit;
  • The absorption is also largely independent of incident angle, which means it can be used for divergent beams too.
Was this FAQ helpful to you? yes no
Close

How do I select the correct wavelength with sensors that have a continual response curve such as silicon photodiode sensors, -LP1/LP2 thermal sensors and 3A-IS sensors?

The sensors with a continual response curve such as the ones listed above come with preset "favorite" wavelengths. If these "favorite" wavelengths do not match the application wavelength you are using they can be changed by performing the instructions below, which are for the Vega meter. For your specific meter, please see the User Manual.

  • While the Vega is off, plug in the head. Switch on the Vega.
  • From the main measurement screen, press "Laser" to select the correct laser wavelength. If you want to save this new wavelength as the startup default, press "Save" before exiting. If the wavelength you want is not among the wavelengths in the six wavelengths listed and you want to change or add a wavelength, see the next step
  • Changing Chosen Wavelengths:
    • From the power measurement screen select "Laser" and enter. Move to the wavelength you wish to change or add. Press the right navigation key.
    • Using the up/down keys to change each number and the right/left keys to move to the next number, key in the desired wavelength. Press the Enter key to exit. If you wish to save this new wavelength as one of the 6 favorite wavelengths, press "Save".

Note: Saving the new wavelength in the Modify screen will not set this wavelength as the default startup wavelength. To do so, you must follow the instructions in Step 2 above.

Was this FAQ helpful to you? yes no
Close

For measuring high average powers, you need to use a high average power sensor (read: big, heavy, water-cooled, expensive). Right?

Well, partly right.

Ophir has for many years had a few sensors that are designed for intermittent use. They are marked by two numbers like 50(150), which means it can measure 50 W continuously, or 150 W for a brief exposure (1.5 minutes in this example). Keeping in mind that power is energy over time, and that it is the total energy absorbed over time that causes a sensor to heat up, it should be possible to expose a sensor to “too high” power but only for a short time, and have the sensor survive the experience. The sensor can treat that short exposure as if it were just one long “single shot” pulse, and measure the energy of that pulse. Divide the energy by the (known) pulse width, and that gives the power during the pulse. (It can’t measure power directly this way, though, since a thermal sensor’s response time to power is itself a few seconds). For example, the moderate-power L40(250)A-LP2-50 has a 10KJ energy scale (several other sensors also have multi kJ scales); to measure power of an 8KW beam, we can fire the laser for 0.5 seconds with the sensor in energy mode, and we’ll measure 4KJ energy in the “pulse”. Dividing that by 0.5 seconds gives the 8KW beam power. Of course we then need to wait for the sensor to cool before repeating, but in some applications that may be perfectly OK
If you have a Juno, Juno+, Centauri or StarBright meter, you can do the above automatically, with any power sensor, using StarBright’s “Pulsed Power” function where you input the pulse duration and the meter will give the readout directly in power.

Was this FAQ helpful to you? yes no
Close

How should I clean my sensor?

First, clean the absorber surface with a tissue, using Umicore #2 Substrate Cleaner, acetone or methanol. Then dry the surface with another tissue. Please note that a few absorbers (Pyro-BB, 10K-W, 15K-W, 16K-W and 30K-W) cannot be cleaned with this method. Instead, simply blow off the dust with clean air or nitrogen. Don't touch these absorbers. Also, HE sensors (such as the 30(150)A-HE-17) should not be cleaned with acetone.
 
Note: These suggestions are made without guarantee. The cleaning process may result in scratching or staining of the surface in some cases and may also change the calibration.

Was this FAQ helpful to you? yes no
Close

Must I use a particular sensor only with the meter it was ordered with?

Ophir meters and sensors are calibrated independently. Each meter has the same sensitivity as the other within about 2 tenths of a percent. Each sensor is calibrated independently of a particular meter with its calibration information contained in the DB15 plug. When the sensor is connected to the meter, the meter reads and interprets this information. Since the accuracy of our sensors is typically +/-3%, the extra 0.2% error that could come from plugging into a different meter is negligible and therefore it does not matter which calibrated meter we use with a particular calibrated sensor.

Was this FAQ helpful to you? yes no
Close

When an accuracy spec is given, what exactly is meant?

The Ophir specification on accuracy is in general 2 sigma standard deviation. This means, for instance, that if we list the accuracy as +/-3%, this means that 95% of the sensors will be within this accuracy and 99% will be within +/-4%. For further information on accuracy see https://www.ophiropt.com/laser-measurement-instruments/laser-power-energy-meters/tutorial/calibration-procedure and https://www.ophiropt.com/laser--measurement/knowledge-center?search=calibration&=SEARCH

Was this FAQ helpful to you? yes no
Close

How do you calculate the power and energy density of a laser beam?

If the power is P and the diameter of the beam is D then the power density is P /(.785 * D2) . If it is a pulsed laser and the energy is E, the repetition rate is R and the diameter is D then the power density is E*R/(.785 * D2), The energy density is E/(.785 * D2). The sensor finder will automatically calculate the power and energy density.

Was this FAQ helpful to you? yes no
Close

Is there a coolant pressure specification for Ophir water-cooled sensors?

Yes. Please reference the chart below:

Minimum Flow Rates for Water-Cooled Sensors

Sensor Recommended flow rate at full power1 Minimum flow rate at full power1 Absolute minimum flow rate pressure drop across sensor (at maximum flow rate) pressure drop across 8 meter of tubing (at maximum flow rate)
liters/min) (liters/min) (liters/min) Bar MPa Bar MPa
L250W 3 3 3 0.3 0.03    
L300W 3 3 3 0.3 0.03    
1000W 6 3 3 0.3 0.03 0.5 0.05
L1500W 6 3.5 3 0.3 0.03 0.5 0.05
L2000W 6 3.5 3 0.6 0.06 0.5 0.05
5000W 8 5 3 0.6 0.06 0.5 0.05
6K-W-200x200 6 6 5 0.5 0.05 0.3 0.03
10K-W 8 8 3 1 0.1 0.5 0.05
15K-W 12 12 3 2 0.2 1 0.1
16K-W 12 12 3 1 0.1 0.8 0.08
30K-W 25 25 6 2 0.2 3 0.3
120K-W 60 60 30 4 0.4 3.5 0.035

You can find a lot more information about the correct use of water-cooled sensors in the article "How to use water cooled Ophir sensors", here.

Was this FAQ helpful to you? yes no
Close

Does the damage threshold depend on power level?

The damage threshold of thermal sensors does depend on the power level and not only the power density because the sensor disc itself gets hotter at high powers. For instance, the damage threshold of the Ophir broadband coating may be 50KW/cm2 at 10 Watts but only 10KW/cm2 at 300W. The Ophir specifications for damage threshold are always given for the highest power of use of a particular sensor, something which is not done by most other manufacturers. This should be taken into account when comparing specifications. The Sensor Finder takes the power level into consideration when calculating damage threshold.

 Watch the 'FAQ: Does damage threshold depend on power level?' video

Was this FAQ helpful to you? yes no
Close

Do I need to recalibrate my instrument? How often must it be recalibrated?

Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.

Was this FAQ helpful to you? yes no
Close

Can a laser measurement depend on the distance from the laser to the sensor?

In theory, if a beam is completely parallel and fits within the aperture of a sensor, then it should make no difference at all what the distance is; it will be the same number of photons (ignoring absorption by the air, which is negligible except in the UV below 250nm). If, nevertheless, you do see such a distance dependence, there could be one of the following effects happening:

  • If you are using a thermal type power sensor, you might actually be measuring heat from the laser itself; when very close to the laser, the thermal sensor might be “feeling” the laser’s own heat. That would not, however, continue to have an effect at more than a few cm distance unless the light source is weak and the heat source is strong.
  • Beam geometry – The beam may not be parallel and may be diverging. Often, the lower intensity wings of the beam have greater divergence rate than the main portion of the beam. These may be missing the sensor's aperture as the distance increases. To check that you'd need to use a profiler, or perhaps a BeamTrack PPS (Power/Position/Size) sensor.
  • If you are measuring pulse energies with a diffuser-based pyroelectric sensor: Some users find that when they start with the sensor right up close to the laser and move it away, the readings drop sharply (typically by some 6%) over the first few cm. This is likely caused by multiple reflections between the diffuser and the laser device, which at the closest distance might be causing an incorrectly high reading. You should back off from the source by at least some 5cm, more if the beam is not too divergent.

Needless to say, it’s also important to be sure to have a steady setup; a sensor held by hand could easily be moved around involuntarily, which could cause partial or complete missing of the sensor’s aperture at increasing distance, particularly for an invisible beam.

Was this FAQ helpful to you? yes no
Close

What is the best water to use in the Water Cooled Sensors?

Corrosion is caused by interactions between the metallic components of the sensor and the cooling water, which may contain a variety of dissolved ions. Many factors affect the risk of corrosion forming, but the most important are the pH and the mixture of ions in the water. For this reason, we recommend using neutral deionized water in a closed circulating system (pH between 6 and 8). Please note that deionized water is usually slightly acidic (pH 5.65) due to absorption of CO2 from the atmosphere. The cooling water can be neutralized by adding 5 ml of a 10 mM solution of NaOH for each liter of water in the cooling system. Commercial additives such as Optishield Plus are also recommended for systems such as ours that have copper and aluminum in them. Optishield has the additional benefit of having biocide to prevent buildup of organic contamination.

To prevent corrosion it is also crucial to not allow standing water to evaporate inside the sensor when it is not in use. When disconnecting a sensor from the cooling system, the water channel should be cleared by blowing compressed air through it.

For those customers still experiencing problems with corrosion, we recommend the new thermal sensor 1000WP-BB-34 which has a special design in which all materials that come into contact with the cooling water are either copper or nonmetallic.

You can find a lot more information about the correct use of water-cooled sensors in the article "How to use water cooled Ophir sensors", here.

Was this FAQ helpful to you? yes no
Close

Once and for all: Is DI (deionized) water good or bad for water-cooled sensors? Does it help prevent corrosion, or does it actually increase the risk?

Many factors affect the risk of corrosion forming, but the two most important are:

  • the mixture of ions in the water
  • the water’s pH

Our current recommendation is to use DI water – but of a neutral pH. DI water is usually slightly acidic; it can be titrated to a neutral pH, using a bit of sodium hydroxide for example. There are also commercial additives that can help prevent corrosion, for instance Optishield Plus. For a more detailed discussion, see the FAQ at https://www.ophiropt.com/laser--measurement/knowledge-center/faq/7805

 

You can find a lot more information about the correct use of water-cooled sensors in the article "How to use water cooled Ophir sensors", here.

Was this FAQ helpful to you? yes no
Close

The sensor I need uses water cooling. Can you recommend a water cooling system?

We don’t supply chillers, nor insist on specific models; the only important thing from our point of view is to simply keep to the requirements specified for the cooling water of the specific model of sensor, such as minimum flow rate at full power, water temperature range, and - more important than the actual water temperature - water temperature stability. The temperature of the water should not be changing by more than 1 deg/min (because changes in water temperature could cause heat flow in the sensor which would be detected as if it were laser power, and cause errors in the reading).

We also have a video on our site at https://www.ophiropt.com/laser-measurement-instruments/laser-power-energy-meters/knowledge-center/water-cooled-sensors-youtube, which discusses various issues and tips about water cooling. There is a short discussion of coolant pressure requirements in our FAQ section at https://www.ophiropt.com/laser--measurement/knowledge-center/faq/2404

Was this FAQ helpful to you? yes no
Close

Videos

LP2 Laser sensors coating: very high damage threshold, very low reflection LP2 Laser sensors coating: very high damage threshold, very low reflection
FAQ: Thermal Sensors for Measuring Low, Medium & High Laser Powers FAQ: Thermal Sensors for Measuring Low, Medium & High Laser Powers FAQ: Thermal Sensors for Measuring Low, Medium & High Laser Powers

In this short “Basics” video we review the use – and selection - of thermal sensors for measuring low, medium and high laser powers.

FAQ: Sensors for Measuring Laser Power FAQ: Sensors for Measuring Laser Power FAQ: Sensors for Measuring Laser Power

In this short “Basics” video we review in general how one goes about measuring laser beam power, so that you’ll have a clear basic understanding of what the different sensor types are, and when you would use which type.

On what factors do Ophir sensor accuracy specs depend? On what factors do Ophir sensor accuracy specs depend? On what factors do Ophir sensor accuracy specs depend?

Ophir's CTO, Dr. Ephraim Greenfield discusses the various factors that contribute to uncertainties in measurement when using Ophir laser power and energy meters. He discusses the various factors and shows how they combine to give the total uncertainty.

Even higher damage threshold…flat spectral response…absorption up to 96%...
These are some of the characteristics of the new “LP2” type laser power sensors from Ophir.
Learn more in this video.

Tutorials

Tutorials and Articles

Laser Power Sensors introduction

As described in the general introduction, the thermopile sensor has a series of bimetallic junctions. A temperature difference between any two junctions causes a voltage to be formed between the two junctions. Since the junctions are in series and the «hot» junctions are always on the inner, hotter side, and the «cold» junctions are on the outer, cooler side, radial heat flow on the disc causes a voltage proportional to the power input. Laser power impinges on the center of the thermopile sensor disk (on the reverse side of the thermopile), flows radially and is cooled on the periphery. The array of thermocouples measures the temperature gradient, which is proportional to the incident or absorbed power. In principle, the reading is not dependent on the ambient temperature since only the temperature difference affects the voltage generated and the voltage difference depends only on the heat flow, not on the ambient temperature.

 阅读更多...

Common Reasons for Thermal Sensor Damage or Out of Tolerance Conditions

This document was created to assist our valued customers in the proper care and maintenance of Ophir thermal laser power sensors. The following information is for reference only. If you have any reason to believe that the sensor is no longer performing within the original specifications,we always recommend that you send it in for repair and/or recalibration by our trained technicians to bring the unit back to the proper NIST traceable standards. We believe that Ophir thermal sensors can be used for many years without any repairs when used with the proper laser optical setup. We hope that this document will enable you to also enjoy the long life and reliable results for which Ophir-Spiricon is known.

 阅读更多...

How to Properly Select a Laser Power or Energy Sensor

The selection of a sensor to accurately measure the power of a laser or energy of a pulsed laser can seem like a simple and easy procedure. However, many times the selection process is limited to choosing a sensor that only meets the range of power or energy to be measured, leaving out several other essential criteria of the laser specifications; that without their consideration, can allow the wrong sensor to be selected, the laser to be measured inaccurately and likely to cause the sensor to fail prematurely.

Watch Our Laser Measurement Video

 阅读更多...

Setting your thermal sensor up in energy mode for the best accuracy and repeatability.

Range Selection Use the lowest range that is larger than the pulse energy to be measured. For example, if you want to measure a 2.7 Joule pulse, use the 3 J range instead of the 30 J range. This will allow for maximum resolution (a 2.700 J reading versus a 2.70 J reading). Threshold Selection For most energy measurements, the default MEDIUM setting is appropriate. If taking measurements in a noisy environment or where there is a high level of background thermal radiation, the instrument may trigger spuriously on the noise or the background radiation. In this case, the user may 阅读更多...

How Much of the Power Sensor Aperture Can My Laser Fill Up?

The entire aperture senses power, so you can use the whole head. That said, a beam in the inner 50% of the surface area (about 70% of the diameter) is specified by Ophir to be uniform within +/-2%. The sensitivity around the edges might be a little less, but generally the sensitivity doesn’t vary by more than +/-2% over the entire aperture.

 阅读更多...

Effect of Ambient Conditions on Laser Measurements

We are often asked about the specified ranges of various ambient conditions (temperature, humidity, etc.) for Ophir instruments. In this article we will clarify the effects of these conditions on laser measurements, so you’ll be able to use your Ophir laser measurement instrument effectively.

 阅读更多...

How to Use Water Cooled Ophir Sensors

Ophir water cooled sensors are designed to measure high powers in a relatively compact package. In order for the sensor to operate properly, the water flow rate, temperature and temperature stability have to be in the right range. For best performance of the sensor, the water flow rate should be the recommended rate. If the user system is not able to reach the recommended rate, then the minimum rate can be used and the sensor will meet spec but there may be some degradation of response time and linearity... 阅读更多...

Ophir Power/Energy Meter Calibration Procedure and Traceability/Error Analysis

This document discusses the interpretation and basis for stated measurement accuracy of Ophir Laser Power/Energy meters.
1. General Discussion
2. Combination of Errors and Total Error
3. Analysis of Power and Energy Calibration Errors
4. Detailed Analysis of Power and Energy Calibration Errors

 阅读更多...

Laser Measurements in Materials Processing: How and When They Absolutely, Positively Must Be Made

19th century British physicist and engineer William Thomson, 1st Baron Kelvin, was the first to say, “If you can’t measure it, you can’t improve it.” When applying this principle to improving laser-based processes, there are a variety of parameters that must be measured. Given the continuously rising power of laser systems in material processing, the requirements for measurement systems are more challenging than ever. Which technologies are available to measure high-power lasers? How often should they be measured? What measurements should be tracked? When this data is collected, what should be done with it? 阅读更多...

How do I know what range, or scale, to set my power/energy meter to? And what happens if I go over range?

Each given range represents one level of gain of an internal amplifier. The electronics, as always, have a limited Dynamic Range. If the measured signal is too low, in other words near the bottom of the range, then it may be lost in the noise and the reading will be inaccurate and noisy. If it’s too high – there may be saturation issues. To give an instrument a usefully wide dynamic range, multiple scales or ranges are used. Switching from range to range can be automatic (“Autorange”), or manual. Autoranging simply starts automatically at the least sensitive range and works its way down the ranges, sampling the signal as it goes, till it finds a range at which the signal is properly detected. Note, by the way, that only in POWER mode is Autoranging available. If we are working in Single Shot Energy mode, there is no Autoranging – simply because when we are measuring a single pulse, the instrument has no opportunity to work its way down the ranges as in Power mode.

 阅读更多...

Types of power / Energy Laser Sensors General Introduction

Power and Single Shot Energy Sensors
Ophir provides two types of power sensors: Photodiode sensors and Thermal sensors. Photodiode sensors are used for low powers from picowatts up to hundreds of milliwatts and as high as 3W. Thermal sensors are for use from fractions of a milliwatt up to thousands of watts.
Thermal sensors can also measure single shot energy at pulse rates not exceeding one pulse every ~5s.

Repetitive Pulse Energy Sensors
For higher pulse rates, Ophir has pyroelectric energy sensors able to measure pulse rates up to tens of KHz. These are described in the energy sensor section, section 1.3.

 阅读更多...

5 Situations Where Laser Performance Measurement is Necessary

Measuring the performance of a laser has possible for a number of years and is accomplished with a variety of techniques. These electronic laser measurement solutions give the laser user more relevant, time-based data that shows trends in laser performance rather than single data points. While these solutions have provided laser users with the ability to present data in a simple and easy to understand manner, the application of the data still seems to be unclear to many laser users. 阅读更多...

Accessories

Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
  • Metric water fittings for all water cooled sensors except 16K-W & 30K-W with quick connection to 10mm plastic tubing. The Metric water fittings are also suitable for the QBH Adapters. Replaces standard fitting connecting to 3/8" tubing (set of 2 each)

  • 3m Cable

    3m Cable

    7E01122A

    * Order only with purchase of a sensor.
    3m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.

  • 5m Cable

    5m Cable

    7E01122B

    * Order only with purchase of a sensor.
    5m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.

  • 10m Cable

    10m Cable

    7E01122C

    * Order only with purchase of a sensor.
    10m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.

  • 12m Cable

    12m Cable

    7E01122D

    * Order only with purchase of a sensor.
    12m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.