5000WP-LP2-50 | Laser Thermal Power Sensors | Power Sensors - Ophir

5000WP-LP2-50

7Z02788
描述: 

5000WP-LP2-50是一款用于高功率密度和长脉冲激光的水冷热电堆功率/能量激光测量探头。孔径为50mm,可以测量20W至5000W的功率。其高损伤阈值LP2涂层可覆盖0.35至2.2μm的光谱范围。该探头专门设计用于非污染水流。探头配有1.5米长的标准电缆,用于连接仪表或PC适配器。

Specification

  • LP2
  • Ø50mm
  • 0.35-2.2µm
  • 20W-5000W
  • N.A.
  • N.A.
  • Ø155 W x 52 D (mm)
  • N.A.
  • 2.5kW/cm²
  • 3 s
  • 0.1J/cm²
  • 130J/cm²
  • 5000W
  • N.A.
  • CE, UKCA, China RoHS
Need help finding the right sensor? Try our 传感器查询

FAQ

I see Ophir has released some new thermal sensors with an absorber called “LP2”. What is it?

The new “LP2” type sensors are specially designed for beams having high power and high power density (and for pulsed beams, high energy density). The LP2 sensors are replacing the equivalent LP1 sensors; as impressive as the LP1 is, the LP2 was developed with the following improvements:

  • Very high damage threshold, for both power density and energy density, for long pulse and CW beams;
  • Spectrally flat; since its absorption remains constant at widely differing wavelengths, this means that sensors based on the LP2 can be used for "white light" or polychromatic beams;
  • Very high level of absorption (as high as 96%, depending on wavelength), meaning much less light is scattered back, which for high power beams is an important benefit;
  • The absorption is also largely independent of incident angle, which means it can be used for divergent beams too.
Was this FAQ helpful to you? yes no
Close

Do I need to recalibrate my instrument? How often must it be recalibrated?

Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.

Was this FAQ helpful to you? yes no
Close

Can a laser measurement depend on the distance from the laser to the sensor?

In theory, if a beam is completely parallel and fits within the aperture of a sensor, then it should make no difference at all what the distance is; it will be the same number of photons (ignoring absorption by the air, which is negligible except in the UV below 250nm). If, nevertheless, you do see such a distance dependence, there could be one of the following effects happening:

  • If you are using a thermal type power sensor, you might actually be measuring heat from the laser itself; when very close to the laser, the thermal sensor might be “feeling” the laser’s own heat. That would not, however, continue to have an effect at more than a few cm distance unless the light source is weak and the heat source is strong.
  • Beam geometry – The beam may not be parallel and may be diverging. Often, the lower intensity wings of the beam have greater divergence rate than the main portion of the beam. These may be missing the sensor's aperture as the distance increases. To check that you'd need to use a profiler, or perhaps a BeamTrack PPS (Power/Position/Size) sensor.
  • If you are measuring pulse energies with a diffuser-based pyroelectric sensor: Some users find that when they start with the sensor right up close to the laser and move it away, the readings drop sharply (typically by some 6%) over the first few cm. This is likely caused by multiple reflections between the diffuser and the laser device, which at the closest distance might be causing an incorrectly high reading. You should back off from the source by at least some 5cm, more if the beam is not too divergent.

Needless to say, it’s also important to be sure to have a steady setup; a sensor held by hand could easily be moved around involuntarily, which could cause partial or complete missing of the sensor’s aperture at increasing distance, particularly for an invisible beam.

Was this FAQ helpful to you? yes no
Close

Is there a coolant pressure specification for Ophir water-cooled sensors?

Yes. Please reference the chart below:

Minimum Flow Rates for Water-Cooled Sensors

Sensor Recommended flow rate at full power1 Minimum flow rate at full power1 Absolute minimum flow rate pressure drop across sensor (at maximum flow rate) pressure drop across 8 meter of tubing (at maximum flow rate)
liters/min) (liters/min) (liters/min) Bar MPa Bar MPa
L250W 3 3 3 0.3 0.03    
L300W 3 3 3 0.3 0.03    
1000W 6 3 3 0.3 0.03 0.5 0.05
L1500W 6 3.5 3 0.3 0.03 0.5 0.05
L2000W 6 3.5 3 0.6 0.06 0.5 0.05
5000W 8 5 3 0.6 0.06 0.5 0.05
6K-W-200x200 6 6 5 0.5 0.05 0.3 0.03
10K-W 8 8 3 1 0.1 0.5 0.05
15K-W 12 12 3 2 0.2 1 0.1
16K-W 12 12 3 1 0.1 0.8 0.08
30K-W 25 25 6 2 0.2 3 0.3
120K-W 60 60 30 4 0.4 3.5 0.035

You can find a lot more information about the correct use of water-cooled sensors in the article "How to use water cooled Ophir sensors", here.

Was this FAQ helpful to you? yes no
Close

What is the best water to use in the Water Cooled Sensors?

Corrosion is caused by interactions between the metallic components of the sensor and the cooling water, which may contain a variety of dissolved ions. Many factors affect the risk of corrosion forming, but the most important are the pH and the mixture of ions in the water. For this reason, we recommend using neutral deionized water in a closed circulating system (pH between 6 and 8). Please note that deionized water is usually slightly acidic (pH 5.65) due to absorption of CO2 from the atmosphere. The cooling water can be neutralized by adding 5 ml of a 10 mM solution of NaOH for each liter of water in the cooling system. Commercial additives such as Optishield Plus are also recommended for systems such as ours that have copper and aluminum in them. Optishield has the additional benefit of having biocide to prevent buildup of organic contamination.

To prevent corrosion it is also crucial to not allow standing water to evaporate inside the sensor when it is not in use. When disconnecting a sensor from the cooling system, the water channel should be cleared by blowing compressed air through it.

For those customers still experiencing problems with corrosion, we recommend the new thermal sensor 1000WP-BB-34 which has a special design in which all materials that come into contact with the cooling water are either copper or nonmetallic.

You can find a lot more information about the correct use of water-cooled sensors in the article "How to use water cooled Ophir sensors", here.

Was this FAQ helpful to you? yes no
Close

Once and for all: Is DI (deionized) water good or bad for water-cooled sensors? Does it help prevent corrosion, or does it actually increase the risk?

Many factors affect the risk of corrosion forming, but the two most important are:

  • the mixture of ions in the water
  • the water’s pH

Our current recommendation is to use DI water – but of a neutral pH. DI water is usually slightly acidic; it can be titrated to a neutral pH, using a bit of sodium hydroxide for example. There are also commercial additives that can help prevent corrosion, for instance Optishield Plus. For a more detailed discussion, see the FAQ at https://www.ophiropt.com/laser--measurement/knowledge-center/faq/7805

 

You can find a lot more information about the correct use of water-cooled sensors in the article "How to use water cooled Ophir sensors", here.

Was this FAQ helpful to you? yes no
Close

The sensor I need uses water cooling. Can you recommend a water cooling system?

We don’t supply chillers, nor insist on specific models; the only important thing from our point of view is to simply keep to the requirements specified for the cooling water of the specific model of sensor, such as minimum flow rate at full power, water temperature range, and - more important than the actual water temperature - water temperature stability. The temperature of the water should not be changing by more than 1 deg/min (because changes in water temperature could cause heat flow in the sensor which would be detected as if it were laser power, and cause errors in the reading).

We also have a video on our site at https://www.ophiropt.com/laser-measurement-instruments/laser-power-energy-meters/knowledge-center/water-cooled-sensors-youtube, which discusses various issues and tips about water cooling. There is a short discussion of coolant pressure requirements in our FAQ section at https://www.ophiropt.com/laser--measurement/knowledge-center/faq/2404

Was this FAQ helpful to you? yes no
Close

Videos

LP2 Laser sensors coating: very high damage threshold, very low reflection LP2 Laser sensors coating: very high damage threshold, very low reflection
Water Cooled Sensors: Things to Look Out For Water Cooled Sensors: Things to Look Out For Water Cooled Sensors: Things to Look Out For

In this video, you will learn about some critical issues you need to consider when using water cooling, such as water temperature, water flow rate, and corrosion prevention.

Even higher damage threshold…flat spectral response…absorption up to 96%...
These are some of the characteristics of the new “LP2” type laser power sensors from Ophir.
Learn more in this video.

Tutorials

Tutorials and Articles

How to Use Water Cooled Ophir Sensors

Ophir water cooled sensors are designed to measure high powers in a relatively compact package. In order for the sensor to operate properly, the water flow rate, temperature and temperature stability have to be in the right range. For best performance of the sensor, the water flow rate should be the recommended rate. If the user system is not able to reach the recommended rate, then the minimum rate can be used and the sensor will meet spec but there may be some degradation of response time and linearity... 阅读更多...

Ophir Power/Energy Meter Calibration Procedure and Traceability/Error Analysis

This document discusses the interpretation and basis for stated measurement accuracy of Ophir Laser Power/Energy meters.
1. General Discussion
2. Combination of Errors and Total Error
3. Analysis of Power and Energy Calibration Errors
4. Detailed Analysis of Power and Energy Calibration Errors

 阅读更多...

Laser Measurements in Materials Processing: How and When They Absolutely, Positively Must Be Made

19th century British physicist and engineer William Thomson, 1st Baron Kelvin, was the first to say, “If you can’t measure it, you can’t improve it.” When applying this principle to improving laser-based processes, there are a variety of parameters that must be measured. Given the continuously rising power of laser systems in material processing, the requirements for measurement systems are more challenging than ever. Which technologies are available to measure high-power lasers? How often should they be measured? What measurements should be tracked? When this data is collected, what should be done with it? 阅读更多...

How do I know what range, or scale, to set my power/energy meter to? And what happens if I go over range?

Each given range represents one level of gain of an internal amplifier. The electronics, as always, have a limited Dynamic Range. If the measured signal is too low, in other words near the bottom of the range, then it may be lost in the noise and the reading will be inaccurate and noisy. If it’s too high – there may be saturation issues. To give an instrument a usefully wide dynamic range, multiple scales or ranges are used. Switching from range to range can be automatic (“Autorange”), or manual. Autoranging simply starts automatically at the least sensitive range and works its way down the ranges, sampling the signal as it goes, till it finds a range at which the signal is properly detected. Note, by the way, that only in POWER mode is Autoranging available. If we are working in Single Shot Energy mode, there is no Autoranging – simply because when we are measuring a single pulse, the instrument has no opportunity to work its way down the ranges as in Power mode.

 阅读更多...

Types of power / Energy Laser Sensors General Introduction

Power and Single Shot Energy Sensors
Ophir provides two types of power sensors: Photodiode sensors and Thermal sensors. Photodiode sensors are used for low powers from picowatts up to hundreds of milliwatts and as high as 3W. Thermal sensors are for use from fractions of a milliwatt up to thousands of watts.
Thermal sensors can also measure single shot energy at pulse rates not exceeding one pulse every ~5s.

Repetitive Pulse Energy Sensors
For higher pulse rates, Ophir has pyroelectric energy sensors able to measure pulse rates up to tens of KHz. These are described in the energy sensor section, section 1.3.

 阅读更多...

5 Situations Where Laser Performance Measurement is Necessary

Measuring the performance of a laser has possible for a number of years and is accomplished with a variety of techniques. These electronic laser measurement solutions give the laser user more relevant, time-based data that shows trends in laser performance rather than single data points. While these solutions have provided laser users with the ability to present data in a simple and easy to understand manner, the application of the data still seems to be unclear to many laser users. 阅读更多...

Accessories

Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
  • Metric water fittings for all water cooled sensors except 16K-W & 30K-W with quick connection to 10mm plastic tubing. The Metric water fittings are also suitable for the QBH Adapters. Replaces standard fitting connecting to 3/8" tubing (set of 2 each)

  • 3m Cable

    3m Cable

    7E01122A

    * Order only with purchase of a sensor.
    3m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.

  • 5m Cable

    5m Cable

    7E01122B

    * Order only with purchase of a sensor.
    5m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.

  • 10m Cable

    10m Cable

    7E01122C

    * Order only with purchase of a sensor.
    10m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.

  • 12m Cable

    12m Cable

    7E01122D

    * Order only with purchase of a sensor.
    12m cable to connect sensor to power meter or interface. Order along with sensor to receive this instead of the standard 1.5m cable.

  • SH to BNC Adapter

    SH to BNC Adapter

    7Z11010

    Allows connection of smart sensors to voltage measuring device for measurement of raw voltage output