Tech Tips
Effect of Ambience Conditions on Laser Measurements

We are often asked about the specified ranges of various ambient conditions (temperature, humidity, etc.) for Ophir instruments. In this article we will clarify the effects of these conditions on laser measurements, so you’ll be able to use your Ophir laser measurement instrument effectively.

Read More»
Damage to sensors or meters is a common occurrence, but can be minimized with proper care.

It is recommended that the sensor be put in a well-ventilated container instead of being put into a tool bag or in an area where other things can contact the sensor surface. The meters too should be put into a container so the screen is protected from objects that might contact and break the meter's screen.

Read More»
Longer Length Sensor Cables

The Ophir sensors are provided with a 1.5m cable between the sensor and the smart head connector. When a longer length cable is needed it can be provided, as long as it is within operational limits. However it is not possible to add an extension to the cable, because that moves the smart head connector away from the meter or interface unit which can degrade the smart head functionality or disable it.

Read More»
How Much of the Power Sensor Aperture Can My Laser Fill Up?

The entire aperture senses power, so you can use the whole head. That said, a beam in the inner 50% of the surface area (about 70% of the diameter) is specified by Ophir to be uniform within +/-2%. The sensitivity around the edges might be a little less, but generally the sensitivity doesn’t vary by more than +/-2% over the entire aperture.

Read More»
Power/Energy Meters: Why Damage Specs are Derated for Larger Beams

For certain pyro sensors with a diffuser, such as the PE50BF-DIF-C P/N 7Z02941, there is a Note (b) that for 10mm beam size the damage threshold specification should be derated by 50%. To explain why the damage specification is derated for a larger beam size, please see the picture illustration below. This has to do with the smaller relative increase in spot size lowering the energy density less on a larger beam than the larger relative increase in spot size with a smaller beam lowering the energy density more.

Read More»
LabVIEW and Ophir Power Meters or PC Interfaces

In order for LabVIEW to work with an Ophir power meter or PC interface, you must install StarLab. LabVIEW does not communicate via the StarLab application. We created a special COM object control for integration intent. You do need to Install StarLab in order to communicate with the device. The installation process of StarLab also installs the necessary USB drivers and registers the OphirLMMeasurement COM object required for LabVIEW (or other user programs) for communication with Ophir power meters and PC interfaces.

If you do not wish to install StarLab on your LabVIEW PC, there is a way. The document available at;OphirLMMeasurement_COM_Object.pdf
describes Ophir's COM object, including how to do it without installing StarLab.

Read More»
Measuring Power of Divergent Beams with Integrating Sphere Sensors

Integrating Sphere
An integrating sphere is used to measure a divergent light source. As shown in the illustration, an integrating sphere has its inner surface coated with a surface that highly reflects (typically 99%) in a scattering, nonspecular way. Thus when a divergent beam hits the walls of the integrating sphere, the light is reflected and scattered many times until the light hitting any place on the walls of the sphere has the same intensity.

Read More»
1. When sending in a Power Meter for repair and/or re-calibration.

Note the settings on your meter and sensor before sending the units in for calibration. To simplify the reintegration of your Ophir measurement instruments back into your system, please record your settings and parameters before sending your devices in for calibration.

During the calibration process occasionally we change the settings on an instrument back to the default. This means that when you receive the equipment back it will likely not start up as you had it. The end-user will see a change in how the meter and/or sensor are behaving. The difference could be as simple as changing the Average function, so the readings now appear less stable.

We try to note the settings and return them to the as-received state, but this isn’t always possible.

Note your settings. It might prevent a call to Customer Service and hours of frustration.

  • Wavelength
  • For measuring Power:
    • Range
    • Average – On or off. If Average is on, note the time –i.e. 3 sec, 10 sec, etc.
  • For measuring Energy:
    • Range
    • Threshold – For thermal sensors, Low, Med, High.
    • For pyroelectric sensors, note the pulse width setting. The new pyroelectric sensors have many other options, such as threshold settings for noisy environments.


While we try to leave your instrument settings as they arrive, this isn’t always the case. Taking a few minutes to note the settings before you send your equipment in for calibration will save time when you reintegrate the devices into your measurement system.

Read More»
Heat Sinking Guidelines

A laser power sensor absorbs laser power while measuring it. If that power is not removed at least as fast as it comes in, the sensor could overheat and fail.

Many of Ophir’s OEM power sensors have a maximum average power specification of “XX watts free standing, YY watts heat sinked.” How does one make sure that a sensor has been provided with proper heat sinking? While we do not provide heat sinks, we do offer some guidelines.

We need to consider two separate issues:

  • Conduction of heat out of the sensor into the heat sink
  • Dissipation of the heat from the heat sink to the air
Read More»
Don’t damage your Ophir power meter by using the wrong power supply.

Ophir power meters use a 12 VDC power supply that supplies 500 mA. This power supply is reverse voltage to most US products, meaning the outside connection is positive (+) and the inside is negative (-). When connecting a power supply to the Ophir power meter, make sure that the power supply is a 12 VDC 500 mA power supply and that the center is positive (+). Some of the newer Ophir power meters like the Vega and Quasar are dual voltage compatible meaning it does not matter if the power supply is center positive or center negative. It still needs to be 12 VDC and 500 mA, but it can be either a center positive or center negative. It is highly recommended that the original power supply included with the equipment be used and if it is missing that Ophir be contacted for a replacement supply.

Read More»
Since introducing the first mobile phone power meter app we have been getting a lot of questions about it. Here are some of the more frequent ones:

 

  • The application does not require “state-of-the-art” phones. It works perfectly OK with 3 year-old HTC Legend (that was not top of the line phone even when it was first announced) and it works well with our test phone which is defined mid-to-low range in today’s phone standards.
  • The software does require android version of 2.3.3 and above. According to latest analysis they account for 65% of current phones (this information comes from google analysis of active phones). Since version 2.3.3 was released in early 2011, almost no new phones are sold with older versions. The contract commitment in the US is usually two years, so most 2.2 and 2.1 phone owners there will probably upgrade to newer phones in the coming months.

To download the Quasar Reader App

Read More»
Tech Tip: How to be Sure of the Calibration of Ophir’s High Power Sensors When They are Calibrated at a Fraction of Full Power

An explanation of how we do this is provided below (A). In addition, a recent check of Ophir’s 5000W head by PTB in Germany shows excellent agreement between our calibration and their standards. The details of the correspondence between our sensor and their standard at powers up to 1400W is included here (B).

A. High Power Measurement Calibration Method and Estimated Accuracy of Models 5000W and 10K-W

Read More»
Setting your thermal sensor up in energy mode for the best accuracy and repeatability.

Range Selection
Use the lowest range that is larger than the pulse energy to be measured. For example, if you want to measure a 2.7 Joule pulse, use the 3 J range instead of the 30 J range. This will allow for maximum resolution (a 2.700 J reading versus a 2.70 J reading).

Threshold Selection
For most energy measurements, the default MEDIUM setting is appropriate. If taking measurements in a noisy environment or where there is a high level of background thermal radiation, the instrument may trigger spuriously on the noise or the background radiation. In this case, the user may select the HIGH threshold setting. This will prevent false triggering and ensure the sensor is measuring the intended pulse. If you are measuring small energies and the unit does not trigger, set the threshold for LOW. LOW threshold may also be used for the best accuracy if the energy measured is less than 10% of the range. For example if measuring less than .3 J in the 3 J range, LOW threshold is often more accurate, and will be more repeatable.

Read More»
Ophir-Spiricon Laser Measurement in Medical Laser Service
By Dan Little, Technical Director, Laser Training Institute, Professional Medical Education Association, INC

The global medical industry incorporates thousands of lasers into its arsenal of treatment tools. Wavelengths from UV to Far-infrared are used for everything from Lasik eye surgery to cosmetic skin resurfacing. Visible wavelengths are used in dermatology and ophthalmology to target selective complementary color chromophores. Laser powers and energies are delivered through a wide range of fiber diameters, articulated arms, focusing handpieces, scanners, micromanipulators and more. With all these variables, medical laser service personnel are faced with multiple measurement obstacles. At the Laser Training Institute (lasertraining.org), with headquarters in Columbus Ohio, we offer a week-long laser service school to medical service personnel. Four times a year, a new class will learn the fundamental concepts of power and energy densities, absorption, optics and most of all how lasers work. With a nice sampling of all the major types of medical lasers, the students learn hands-on calibration, alignment and multiple service skills.

Read More»
Can Ophir pyroelectric sensors measure single shot energy or energy at very low repetition rates?

Ophir pyroelectric sensors can measure energy at very low repetition rates, what is called “single shot” energy as well as at various repetition rates all the way up to the maximum in the specification for such sensors. There seems to be a misunderstanding among users that pyroelectric sensors cannot measure single shot energy. This probably comes about since thermal sensors can only measure at very low repetition rates (~0.2Hz), then it is assumed that the converse applies to pyroelectric sensors, i.e. that they only can only measure at faster repetition rates, but this is not true.

Read More»
Measuring Unstable Laser Power

What is the best way to measure the power of a laser that is, unfortunately, not stable, where the power is drifting upwards or downwards. I am making a measurement taking readings by hand and logging to computer?

The best way to measure in such a condition is to take statistics of a number of readings. In order for the sample to be truly random, you should a given intervals look at the reading and take it down as seen with no attempt to wait for it to stabilize or reach a "better" value.

Read More»
When you log energy from multiple sensors in parallel using StarLab, the time stamps in the various channels are synchronized with each other.

If, for example, you need to measure energy at various points along an optical setup in order to characterize each stage of your system, you can place a sensor at each location along the way and connect the sensors in parallel to, say, a multi-channel Pulsar, and log the data using the StarLab application. So long as you open all channels in a single window and log from that window, all the channels will be synchronized with the same zero point. Knowing that, you can rely on the time stamps to tell you which pulse in each channel corresponds to which pulse in the other channels.

Read More»
Fiber Adapter accessory

When using a Fiber Adapter accessory together with one of Ophir's sensors, it is important to be aware of the power/energy density that is going to reach the sensor's surface.In most cases, the fiber adapter locates the fiber tip far enough away from the absorber surface that the spot diameter on the surface will be large, and problems of damage avoided. However, that is not always a certainty! For example, when using a sensor from the PD300 series, especially with filter IN, this distance could be quite small. A power level that is within spec limits could still result in a power density on the filter that is too high, and the result could be damage to the filter. In most cases, as noted, this is not a concern, but in cases where any one parameter seems like it is going to be near a limit, a quick sanity check is a good idea.

Read More»
Tech Tip: A customer has a laser that is only on for about 0.3s and then off for 2s repetitively and wants to measure the peak power using a thermal sensor. Our fastest thermal sensor has a response time of about 0.8s so this is not fast enough for the reading to stabilize and get a reliable reading.
by Ephraim Greenfield, Manager, R&D, Ophir Optronics
Instead of measuring power we measure total energy and the meter is fast enough to read out the energy and be ready for the next pulse 2.3s later. The accuracy of this method is better than +/-1%
Read More»
Tech Tip: Calibration Method and Estimated Accuracy of Ophir High Power Meters: 1000W/1500W/5000W/8000W
by Ephraim Greenfield, Manager, R&D, Ophir Optronics
Ophir high power laser power meters are calibrated using relatively low power lasers (~ 120W CO2 and ~200W YAG). Using such a low power laser to calibrate the instrument vs. the high power at which it is used raises the question of the accuracy of calibration, hence the following analysis.

The high power sensors work on the thermopile principle, where the heat flows radially in the absorber disk causing a temperature difference between the hot and cold junctions of the thermopile which in turn causes a voltage difference across the thermopile.
Read More»
Tech Tip: What Are the Special Requirements in Measurement of 193nm Radiation?
193nm excimer laser radiation needs special precautions in measurement because of its strong interaction with ordinary matter. This radiation is absorbed by ordinary air and water vapor in the air so that the intensity measured can vary by 1% per cm.
Read More»