LD 8900HDR Far-Field Profiler
LED Profiler

Mode Field Diameter

 

Highest dynamic range Goniometric Radiometer and Fiber Optic Analyser
Measurement Parameters

  • Laser Diodes on wafer, chip or package
  • Effective Area (Aeff):
  • Numerical Aperture (NA):
  • Direct measurement of Waveguides and other devices
  • 3D and Single Axis
  • 180 degree field of view (FOV)
  • LED Profiler configuration with 260 degree FOV
Description
Specifications
Ordering Info
Catalog/Manual

Download Data Sheet.

The photonics community needs a method for very high dynamic range measurements of optical fiber, waveguide and other optical components. Advances in DWDM technology, deployment of standard fibers, and the development of specialty fibers all require more accurate characterization. These high-dynamic-range measurements of irradiant light flux have been very difficult to obtain. Photon has the solution: our Goniometric Radiometer, the LD 8900HDR Far-Field Profiler.

Measurement Parameters
The LD 8900HDR Far-Field Profiler uses goniometric methodology to measure true 3-dimensional profiles of optical components with up to 93 dB dynamic range. All measurements are taken with the emitting fiber, planar waveguide, laser diode or LED inserted directly into the center of the instrument's radial measurement area. For optical fiber, the far-field distribution pattern is acquired without bending the fiber. Bending the fiber can cause analysis and reporting inaccuracies. Parameters measured for optical fiber are:

  • Mode Field Diameter (MFD) – used to evaluate losses due to mismatch at connections of fibers and fiber components.
  • Effective Area (Aeff) – used to assess non-linear effects
  • Numerical Aperture (NA) – a measure of the angular width of the fiber output. NA is equal to the sine of the half angle (sin θ/2) at aspecified percentage of the peak value.
The Model LD 8900HDR analysis software calculates these optical fiber parameters from the far-field profile, the method described in the TIA/EIA FOTP-191 standard for MFD requiring a minimum of 50dB for proper measurement. The TIA/EIA FOTP-132 standard for A requires a minimum of 50dB for proper measurements as well.

This wider dynamic range provides greater signal-to-noise detail in the tails of the far-field pattern. Real-time data gathering and reporting of sample optical fibers or other sources is valuable for manufacturing and quality assurance evaluation.

Hardware and Computer
The LD 8900HDR is a modular system that allows either one-dimensional or three-dimensional measurements.
The 3D module is easily exchanged with the 1D hardware, and the system can be purchased with the 1D hardware, the 3D hardware or both. There are specially designed fiber mounting clips available for both the 1D and 3D configurations with a specially modified fiber cleaver and clip that guarantees proper fiber alignment. In addition, custom mounting devices are available for high dynamic range measurements of waveguides, laser diodes and LEDs.

The controller comprises the PC microprocessor, data acquisition card, scan-control card and LD 8900HDR acquisition and analysis software. A reference calibration standard (NIST traceable), 1310nm, 0dBm LD Light Source is also available. This is the measured reference at the time of original manufacture that can be used for unit verification by the user at a later date. For high dynamic range measurements in the visible spectrum, a silicon detector can be substituted with a spectral range of 320-1100nm and even higher 103dB dynamic range. Maximum light source unit size for direct measurements is approximately 12.7mm by 12.7mm.

In the 3D configuration the LD 8900HDR includes the 3D Motion Control Module and is capable of full 3D measurements with up to 200 azimuthal planes through the source.

  
LD 8900HDR Specifications
Specifications 1D Single-Scan Data 3D Multi-Scan Data
Parameters measured MFD; NA; Aeff MFD; NA; Aeff; Angular Width; Angular Centroid; Angular Aspect Ratios, Angular Peak Position, Relative Integrated Power, Relative Power in user-specified cone angles (NA), Cone Angles (NA) in user-specified % Relative Power
Accuracy   MFD: ±0.5% (for nominal 9mm diameter fiber)
    NA: <±1%
    Aeff: <±1%
Scan range   ±90°
Scan radius   13.26(cm); 5.22(in.)
Scan time   <20 seconds
Dynamic range   >93dB InGaAs (103dB Si); input power and distribution dependent; 0 dBm (1mW) minimum for >50dB dynamic range
Spatial sampling resolution   0.055°
Spectral range   InGaAs detector: 800-1700nm / Silicon available (350-800nm)
Software   Windows 2000/XP Professional; ActiveX Automation server capability
Software views Single scan rectangular profile plot; Fields for operator name, fiber type, date, and user comments Polar profile plot - single or dual orthogonal scans; Rectangular profile plot - single or dual orthogonal scans 3D Polar view; 3D Rectangular View; 2D Topographic View; Beam parameter statistics; Time statistics charts; Relative angular power distribution; Notes window for appending user comments
Scan unit dimensions   267 H x 305 W x 362 L (mm); 10.5 x 12 x 14.25 (in.)
Scan unit weight   ~13.6kg; 30lbs.
Maximum light source unit size for direct measurements   ~12.7 × 12.7(mm); 0.5 × 0.5(in.)
      Item Description P/N  
     
  LD8900RHD Very high dynamic range Goniometric Radiometer designed to measure Mode Field Diameter (MFD) and Effective Area (Aeff) of single mode fiber. PH00198  
         
  CC Photon Fiber Clip and Clip Cartridge. The Photon Fiber Clip mounts in a modified Fitel Furukawa Cleaver.
The fiber is cleaved and positioned simultaneously. Cartridge inserts into the 1D Cartridge.
PH00202  
         
  FFC Fiber Cleaver for use with the Photon Fiber Clip. PH00203  
         
  REF Standard reference cartridge Includes a Photon Standard SMF-28 Reference Cartridge and a 1310nm-0dBm LD Light Source. PH00204  
         
  BFC Bare fiber connector for LD8900HDR Cartridge has a v-groove and clamp for holding fiber. Fiber is positioned manually to a visual reference position. Intended for use with specialty fibers that canno. PH00205  
         
  3D 3-D scan module for HDR Interchangeable accessory adds full 3-dimensional measurement capability with up to 200 cuts through the fiber. Includes a 3D Scan Module and Motion Controller. PH00206  
         
  CFP Connectorized fiber mount for HDR Adapter plate for mounting connectorized fibers. Standard ST, FC, and SC connectors are held by the ferrule tip. Call for other types. PH00208  
         
  Case Rugged shipping case for LD8900HDR PH00209  
         
  MfgSoft Software interface designed for manufacturing. Simple interface screen shows the high dynamic range 1D profile and reports MFD, Aeff, and NA. Includes fields for time stamps, serial numbers, and notes. PH00210  
         
  1W Mount 1-D waveguide mount Waveguide mount for use with 1D Interface Plate. PH00215  
         
  1CustomFxt Custom mnt for 1-D vers of HDR/LED Prof Custom designed fixtures for mounting novel sources for 1D scans. PH00217  
         
  LED Profiler A goniometric profiler designed to rapidly measure true 3-dimensional profiles of LEDs providing a ±130 degree measurement around the source, with 0.05-degree resolution. PH00199  
         
  LED profiler w/3180 Same as above but includes state-of-the- art PC running Microsoft Windows operating system with Photon LED Software installed. Configuration required for international sales. PH00200  
         
  3D(LED) 3-D scan module for LED Profiler. PH00207  
         
  3LED Mount Standard LED mount for use with packaged LEDs. Price dependent on configuration. (Must be quoted for specific device.) PH00211  
         
  3CustomFxt Photon can provide custom and semi custom device mounts for LED devices, including TE-cooled vacuum chucks for bare chip testing. Mounts are designed to take advantage of the 260 degree field-of-view available in the LED Profiler. Mounting hardware that is compatible with automated test operations can also be designed and supplied by Photon. PH00214  
         
  UV Allows the use of the LED Profiler with sources emitting at wavelengths below 380nm. PH00212  
         
  PO Pulsed Operation for LED Profiler. PH00213  
         
  Beam Profilers Beam Profilers Catalog
107 pages (8.12 MB )
  BeamGage LD 8900HDR Manual
99 pages (2.45 MB )